This series is a kind of high-performance and high-intelligence equipment for sterilizing,leak detecting and cleansing, which adopts colorful touching screen and PLC ampoule, oral medicine, small infusion bottle, etc. After sterlization, by means of waster is assured to be 100%. Finally washing treatment is carried out sith clean water to enaure cleanly and uncontaminated surface of bottle.
Model |
Overall Dimensions L*W*H(mm) Single/Double Door | Chamber Size L*W*H(mm) | Steam (kg/cycle) | Compressed Air (m3/cycle) |
Cooling Water (kg/cycle) |
Sterilizing Car (pcs) |
Power (KW) |
Net Weight (kg)
Double door |
AM-0.3 | 1195*1220*1720 | 1000*600*600 | 55 | 0.5 | 300 | non | 3 | 760 |
AM-0.6 | 1245*1300*1880 | 1050*180*850 | 75 | 0.8 | 400 | non | 3 | 1100 |
AM-1.2 | 1695*1370*1960 | 1500*750*1100 | 85 | 1.5 | 800 | 2 | 4 | 1900 |
AM-1.8 | 2445*1370*1960 | 2250*750*1100 | 100 | 2.0 | 1200 | 3 | 4 | 2450 |
AM-2.5 | 3195*1370*1960 | 3000*750*1100 | 180 | 3.0 | 1500 | 4 | 6 | 2800 |
AM-5.0 | 3635*1800*2200 | 3400*1000*1500 | 300 | 5.5 | 2500 | 4 | 6 | 45 |
Structure & Capability & Characteristics
Fine working case body and chamber:Rectangle case body with the strengthen tendon is designed and made according to the 《Steel Pressure Vessel》 standard and 《Pressure Vessel Technical Detection Regulation 》.
The inside layer is welded by good quality acid-resisting stainless steel 304. The out wll and strengthen tendon is made of good quality carbon steel. The chamber inside surface is mechanical polished, smooth, good outlook and anticorrosion.
Patent door lock device:
Cylinder-driving device is the pioneer in the same occupation in domestic market. It is easy to operate without maintain. The reliable mechanical electric interlock make ensure the safe operating.
Seal trip:
The gasket ring is made of good quality silicone rubber which can resist 200℃ and higher temperature. The unique inner gas-filled structure ensure the seal tightness.
Reliable stabile micro-computer automatical control and advanced function for recording, printing and storing, etc.
As the main system controller, touch-screen+PLC is selected worldwide well-known brand and reliable. In the matter of data managing, the real-time print or CF store can be choosed.
Elegant decorate cover:
The cover is made of wiring-drawn stainless steel, good outlook, serviceable and easy to clean.
Unique spraying cooling device:
The spiral spraying mouths from Spraying Systems Co. USA are fixed in this spraying system. The sprayed cooling water is distributed thickly dotted in the chamber without dead corner that can make the ampoules cooling uniformly.
The GMP confirmation access is fixed which confirm the local temperature and sterilizing result.
Reliable Leak Detection
Vacuum leak testing:
The bad quality ampoules or the ampoules in a certain vacuum condition will demolish or leak, so it is easy to check out the unqualified ampoules after vacuum testing.
Dyestuff leak testing,First, pour the dyestuff water over the level of the sterilizing goods into the sterilizing chamber.
Then rise the pressure to make the dyestuff water into the ampoule. After the water output, the bad ampoules which have dyestuff remains can be check out.
Main Parameters and Utilities
Design Pressure | 0.245Mpa | Vacuum | -0.095Mpa |
Working Pressure | 0.22Mpa | Steam Pressure | 0.4-0.6Mpa |
Design Temperature | 139℃ | Water Pressure | |